Shoulder Simulator

Cable Actuated Shoulder Simulator

Technical Specifications:

  • Joint Reaction Force: 2 kN max.
  • Abduction: 60°+
  • Testfrequency: ≤50 Hz 
  • Testing space: ~ 2500 × 800 × 1500 mm (L × B × H)
  • Drive System: Electrodynamic rotary motors
  • Testing types: Investigation of rotator cuff ruptures and their compensation or implant performance during abduction movement
  • Data acquisition & control: Control via LabVIEW and MATLAB with integration of external sensors; motion capture is recorded by Optitrack
  • Available load sensors: Tension sensors (4x450 N, 4x 2.5 kN), and 6DoF compression 
  • sensor (2 kN;  500N Shear)
  • Can be combined with other measurement systems: DIC, MoCap, IMU
  • Number of available systems: 1

Virtual Lab Tour